Fast Current Transformer (FCT) is used for non-destructive measurement of pulsed beam current. It is a passive, in-flange FCT. It consists of stainless steel CF150 conflate flanges, a ferrite core and an Alumina ceramic brake. A BNC connector is provided for measurement. It can be used to measure pulse beam currents of duration up to 10 µs, has a sensitivity 3 V/A and droop is less than 5%/ µs. It is UHV compatible.
Fast Current Transformer (FCT) is used for non-destructive measurement of pulsed beam current. It is a passive, in-flange FCT. It consists of stainless steel CF150 conflate flanges, a ferrite core and an Alumina ceramic brake. A BNC connector is provided for measurement. This FCT can be used to measure pulse beam currents of duration up to 10 µs, has a sensitivity 3 V/A and droop is less than 5%/ µs. It is UHV compatible.
APPD, BARC has developed an In-flange, passive FCT consisting of SS304 conflate flanges, a ferrite core and Alumina ceramic brake. A BNC connector is provided for the measurement. FCT is UHV compatible and can be used to measure pulsed beam currents having pulse width up to 10µs. The Sensitivity is 3V/A and Droop is less than 5%/µs.
Physical Parameters
1 | Size | OD: 203 mm, ID: 52 mm, length: 44 mm |
---|---|---|
2 | End connection | CF150 conflate flanges |
3 | Flange Material | SS304 |
Electrical Parameters
1 | Sensitivity | 3V/A (when terminated with 50 Ω resistor) |
---|---|---|
2 | Voltage Droop | : ≤ 5 %/ µs (linear up to 8 µs pulse width) |
3 | Maximum Pulse width | : ~10 µs |
4 | Maximum Peak Beam Current | : 500 mA |
Infrastructure
A laboratory having following equipments :
Statutory Requirements
RAW MATERIALS
MANPOWER